Total Titratable Alkalinity System Including pH and Temperature

The HI 84431 is an automatic titrator designed for easy, fast and accurate analysis of Total Titratable Alkalinity in water. Alkalinity in water is its acid neutralizing ability.

( Manufacturer )
$889.99

Ask a question about this product

Total Titratable Alkalinity System Including pH and Temperature monitoring (HI 84431-01)

hi_84431_sm.jpgThe HI 84431 is an automatic titrator designed for easy, fast andaccurate analysis of Total Titratable Alkalinity in water.Alkalinity in water is its acid neutralizing ability.

Potentiometricendpoint determination, peristaltic titrant delivery and integratedmagnetic stirring systems eliminate the error and technique dependentresultsassociated with manual titrations.

The accuracy of the instrument is ensured by performing aperistaltic pump calibration using the provided HANNA standard. The HI84431 endpointdetermination algorithm analyzes the pH vs. volume curve to determinethe exact pH endpoint and performs the necessary calculations. Theresultsare displayed in mg/L CaCO3 or meq/L CaCO3 units on the graphic display.

titration_1.jpgAn intuitive interface makes the instrument simple to use. Adedicated HELP key guides the user through set-up and calibrationsequences, reportsinstrument status and aids in troubleshooting.

HI 84431 features include: log on demand for up to 100 samples (50pH measurements, 50 titration results), GLP compliance and PC interfacevia USB using the HI 92000 windows software (sold seperately).

titration_2.jpgThe HI 84431 determines the total titratable alkalinity in water bypotentiometric titration. As titrant is slowly added to the samplesolution thepH and temperature are carefully monitored. The software analyzes theresulting pH curve and determines the volume of titrant required toreachthe endpoint.

Titrations are conducted using the low range reagent (10 to 500 mg/L as CaCO3) or the high range reagent (400 to 4000 mg/L as CaCO3).

This complete system is supplied with the following accessories. Replacement accessories are available on request. 

  • HI 1131B pH electrode,
  • HI 7071 filling solution (30 mL),
  • HI 7662-M temperature probe,
  • HI 84431-50 titrant low range (100 mL),
  • HI 84431-51 titrant high range (100 mL),
  • HI 84431-55M pump calibration solution (230 mL),
  • HI 7061 cleaning solution,
  • HI 7004M pH 4.01 buffer solution (230 mL),
  • HI 70083M pH 8.30 buffer solution (230 mL),
  • HI 7010M pH 10.01 buffer solution (230 mL),
  • 100 mL beakers (2),
  • tube set with dispensing tip,
  • Stir bars (medium 2 pcs.),
  • 12 Vdc power adapter,
  • Instruction manual and quick reference guide. 
Range Titratable alkalinity (Low range) mg/L : 10.0 - 500.0 mg/L as CaCO3
meq/L : 0.2 - 10.0 meq/L as CaCO3
Range Titratable alkalinity(High range) mg/L : 400 - 4000 mg/L as CaCO3
meq/L : 8 - 80 meq/L as CaCO3
Resolution Titratable alkalinity(Low range) 0.1 mg/L
0.1 meq/L
Resolution Titratable alkalinity(High range) 1mg/L
1meq/L
Accuracy   5% of reading
Titration method   Acid-base titration (total alkalinity / strong acidity)
Principle End point titration 4.5 pH / 8.3 pH
Pump debit   0.5 mL/min
Stirring speed   600 rpm
Log data   Up to 50 samples
Range pH -2.0 to 16.0 pH / -2.00 to 16.00 pH
Resolution pH 0.1 pH / 0.01 pH
Accuracy pH ± 0.01 pH
Calibration   1, 2 or 3 calibration points; 3 available buffers (4.01; 8.30; 10.01)
Temperature compensation   manual or automatic from -20 to 120 °C (-4 to 248 °F)
Log data   Up to 50 samples
Range Temperature -20.0 to 120.0 °C (-4.0 to 248.0 °F)
Resolution Temperature 0.1 °C
Accuracy Temperature ±0.4 °C without probe error
Electrode   HI 1131B (included)
Temperature Probe   HI 7662-M (included)
Environment   0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply   12 Vdc power adapter
Dimensions   208 x 214 x 163 mm (8.2x8.4x6.4") (with beaker)
Weight   2200 g (77 oz.)

 Unit Ship Weight: 7Lbs