Total Titratable Acidity System Including pH and Temperature monitoring (HI 84430-01)
The HI 84430 is an automatic titrator designed for water analysis that is easy, fast andaccurate. Potentiometricendpointdetermination, peristaltic titrant delivery and integrated magneticstirring systems eliminate the error and technique dependent resultsassociatedwith manual titrations.
The accuracy of the instrument is ensured by performing aperistaltic pump calibration using the provided HANNA standard. The HI84430 endpointdetermination algorithm analyzes the pH vs. volume curve to determinethe exact pH endpoint and performs the necessary calculations. Theresultsare displayed in mg/L CaCO3 or meq/L CaCO3 units on the graphic display.
An intuitive interface makes the instrument simple to use. Adedicated HELP key guides the user through set-up and calibrationsequences, reportsinstrument status and aids in troubleshooting.
This titration method is not subject to the same interferences as a colormetric technique and the peristaltic pump assure the best performance and repeatability possible.
HI 84430 features include: log on demand for up to 100 samples (50pH measurements, 50 titration results), GLP compliance and PC interfacevia USB to use with the HI92000 windows software (Sold Seperately).
Titrations are conductedusing the low range reagent (15 to 500 mg/L as CaCO3) or the high rangereagent (400 to 4000 mg/L as CaCO3).
This complete system is supplied with the following accessories. Replacement accessories are available on request.
- HI 1131B pH electrode,
- HI 7071 filling solution (30 mL),
- HI 7662-M temperature probe,
- HI 84430-50 titrant low range (100 mL),
- HI 84430-51 titrant high range (100 mL),
- HI 84430-55 pump calibration solution (230 mL),
- HI 84430 additional reagent (30 mL),
- HI 7061 cleaning solution,
- HI 7001M pH 1.68 buffer solution (230 mL),
- HI 7004M pH 4.01 buffer solution (230 mL),
- HI 70083M pH 8.30 buffer solution (230 mL),
- 100 mL beakers (2), tube set with dispensing tip,
- stir bars (medium 2 pcs.),
- 12 Vdc power adapter,
- instruction manual, and quick reference guide.
Specifications:
| Range |
Titratable acidity (Low range) |
mg/L : 15.0 - 500.0 mg/L as CaCO3 meq/L : 0.3 - 10.0 meq/L as CaCO3 |
| Range |
Titratable acidity (High range) |
mg/L : 400 - 4000 mg/L as CaCO3 meq/L : 8 - 80 meq/L as CaCO3 |
| Resolution |
Titratable acidity (Low range) |
0.1 mg/L 0.1 meq/L |
| Resolution |
Titratable acidity (High range) |
1mg/L 1meq/L |
| Accuracy |
|
5% of reading |
| Titration method |
|
Acid-base titration (total acidity / strong acidity) |
| Principle |
End point titration |
8.3 pH / 3.7 pH |
| Pump debit |
|
0.5 mL/min |
| Stirring speed |
|
600 rpm |
| Log data |
|
Up to 50 samples |
| Range |
pH |
-2.0 to 16.0 pH / -2.00 to 16.00 pH |
| Resolution |
pH |
0.1 pH / 0.01 pH |
| Accuracy |
pH |
± 0.01 pH |
| Calibration |
|
1, 2 or 3 calibration points; 3 available buffers (1.68; 4.01; 8.30) |
| Temperature compensation |
|
manual or automatic from -20 to 120 °C (-4 to 248 °F) |
| Log data |
|
Up to 50 samples |
| Range |
Temperature |
-20.0 to 120.0 °C (-4.0 to 248.0 °F) |
| Resolution |
Temperature |
0.1 °C |
| Accuracy |
Temperature |
±0.4 °C without probe error |
| Electrode |
|
HI 1131B (included) |
| Temperature Probe |
|
HI 7662-M (included) |
| Environment |
|
0 to 50 °C (32 to 122 °F); max 95% RH non-condensing |
| Power supply |
|
12 Vdc power adapter |
| Dimensions |
|
208 x 214 x 163 mm (8.2x8.4x6.4") (with beaker) |
| Weight |
|
2200 g (77 oz.) |
Unit Ship Weight: 7Lbs
|