PrintE-mail

Extech EasyView Light Meter, 999,900 Lux With Memory - EA33

This device measures up to 99,990 Fc and 999,990 Lux and stores up to 50 readings for easy recall. Includes ripple function to exlude effect of stray light.

( Manufacturer )
$259.99

Ask a question about this product

Extech EasyView Light Meter, 999,900 Lux With Memory - EA33

  • EA33_SFeatures:

    • Large LCD display with advanced illuminance functions and features
    • Wide measurement range to 99,990Fc (999,900 Lux) with resolution of 0.001Fc and 0.01Lux
    • Luminous intensity (candela) calculations
    • Store and recall up to 50 measurements, includes relative or real time clock stamp
    • Ripple function excludes the effect of stray light from the primary light source measurement
    • Timed-Hold, Relative in absolute value or % deviation, and Comparator function with high low alarms.

    The Extech EA33 light meter features a wide measurement range to 999,900 Lux with a resolution of 0.01Lux. The ripple function allows you to exclude the effect of stray light and maintain high precision within 3%. With the Extech EA33 light meter's built-in memory, you can store up to 50 measurements at a time.

    The Extech EA33 light meter features an easy-to-read display, which is also cosine and color corrected. The timed-hold function is relative in absolute value or % deviation.

    The handheld Extech EA33 light meter, weighing only 11.29 oz., comes complete with built-in stand, light sensor, and protective cover with 36" (0.9m) cable, protective holster, 6 AAA batteries, and carrying case.

    Specifications

    Units Ranges and Resolution Accuracy

    99.99, 999.9, 9999 ± (3% rdg + 0.5%FS)

    99990 ± (3% rdg + 5 digits)

    Lux

    • 999900 ± (3% rdg + 0.5%FS)
    • 9.999, 99.99, 999.9 ± (3% rdg + 0.5%FS)
    • 9999 ± (3% rdg + 5 digits)

    Foot candles

    99990 ± (3% rdg + 0.5%FS)

    Note: Calibrated to a standard incandescent lamp at color temperature of 2856K

    • Display Multi-function LCD with a maximum display of 999900
    • Ranging Autoranging over five ranges
    • Over range indication LCD displays ‘OL’
    • Spectral response CIE photopic (CIE human eye response curve)
    • Spectral accuracy Vλ function (f’1 ≤6%)
    • Cosine response f’2 ≤2%; Cosine corrected for angular incidence of light
    • Measurement Repeatability ±2%
    • Temperature coefficient ±0.1% per oC
    • Sampling rate 5 times per second
    • Photo detector Silicon photo-diode with spectral response filter
    • Operating conditions Temperature: 32 to 104oF (0 to 40oC); Humidity: < 80%RH
    • Storage conditions Temperature: 14 to 140oF (-10 to 50oC); Humidity: < 80%RH
    • Meter Dimensions 5.91 x 2.8 x 1.4" (150 x 72 x 35mm)
    • Detector Dimensions 3.6 x 2.4 x 1.5” (92 x 60 x 29mm)
    • Weight Approx. 11.3 oz. (320g) with battery
    • Sensor lead length 5’ (150cm)
    • Low battery indication Battery symbol appears on the LCD
    • Power supply Six (6) ‘AAA’ cells
    • Battery life 200 hours (approx.)

    Unit ship weight: 3Lbs